Albuterol samples collected in the inhalation filter, nebulizer, T-piece, and corrugated tubing were eluted with distilled water and analyzed with a spectrophotometer. RESULTS: The inhaled drug, as a percentage of total dose in both lung models, was 5.1-7.5%, without statistical significance among the 3 modes. Median nebulization times for IIM, CM, and EIM were 38.9, 14.3, and 17.7 min, respectively, and nebulization time for the 3 modes significantly differed (P < .001). The inhaled drug mass for the 3 modes with the adult lung model was similar to that with the pediatric lung model (7.39 ؎ 0.76 vs 6.27 ؎ 0.69%, P ؍ .77). CONCLUSIONS: Aerosol drug delivery with a jet nebulizer placed proximal to the ventilator was not dependent on nebulization mode during simulated pediatric and adult conventional mechanical ventilation. Use of expiratory intermittent mode and continuous nebulization should be considered to reduce treatment time.