“…An agreement study between subjective refraction and objective refraction measured with AR (KR-8800, Topcon) and HS (Wasca, Zeiss) showed a mean bias of −0.3D in eyes with monofocal IOLs and with 5% of the measured eyes with a difference > 0.75D. 15 This bias is consistent with that obtained using SS (OPD Scan III) for a central measurement area of 2.6 mm, however a higher myopic error has been reported for larger measurement areas, also in enhanced monofocal IOLs (Tecnis Eyhance). 21 , 22 Less bias has been reported by other authors, such as around −0.14D for AR (KR-8800, Topcon) and RT (iTrace, Tracey), which increases until −0.39D for HS (LADARWave, Alcon).…”