Objective: The aim of this work was to investigate the main dosimetric characteristics and the performance of an A26 Exradin ionization microchamber (A26 IC) and a W1 Exradin plastic scintillation detector (W1 PSD) in small photon beam dosimetry for treatment planning system commissioning and quality assurance programme. Methods: Detector characterization measurements (short-term stability, dose linearity, angular dependence and energy dependence) were performed in water for field sizes up to 10 3 10 cm 2 . Polarity effect (P pol ) was examined for the A26 IC. The behaviour of the detectors in small field relative dosimetry [percentage depth dose, dose profiles often called the off-axis ratio and output factors (OFs)] was investigated for field sizes ranging from 1 3 1 to 3 3 3 cm 2 . Results: Results were compared with those obtained with other detectors we already use for small photon beam dosimetry. A26 IC and W1 PSD showed a linear dose response. While the A26 IC showed no energy dependence, the W1 PSD showed energy dependence within 2%; no angular dependence was registered. P pol values for A26 IC were below 0.9% (0.5% for field size .2 3 2 cm 2 ). A26 IC and W1 PSD depth-dose curves and lateral profiles agreed with those obtained with an EDGE diode. No differences were observed among the detectors in OF measurement for field sizes larger than 1 3 1 cm 2 , with average differences ,1%. For field sizes ,1 3 1 cm 2 , the effective volume of ionization chamber and non-water equivalence of EDGE diode become significant. A26 IC OF values were significantly lower than EDGE diode and W1 PSD values, with percentage differences of about 223 and 213% for the smallest field, respectively. W1 PSD OF values lay between ion chambers and diode values, with a maximum percentage difference of about 210% with respect to the EDGE diode, for a 6 3 6-mm 2 field size. Conclusion: The results of our investigation confirm that A26 IC and W1 PSD could play an important role in small field relative dosimetry. Advances in knowledge: Dosimetric characteristics of Exradin A26 ionization microchamber and W1 plastic scintillation detector for small field dosimetry.