2023
DOI: 10.1038/s41598-023-47198-2
|View full text |Cite
|
Sign up to set email alerts
|

Analysis of flow rate of continuous bladder irrigation according to the height of the irrigation infusion set

Boeun Yang,
Jeongwon Han

Abstract: This is a control volume analysis to examine the flow rate of irrigation fluid according to the size of indwelling catheter and the height of the fluid bag in consideration of the temperature of irrigation fluid and intra-bladder pressure during continuous bladder irrigation. In case of minimum bladder pressure with room temperature, the flow rates were − 0.045 to 0.993 cc/sec for 18Fr, − 0.053 to 1.176 cc/sec for 20Fr, − 0.055 to 1.227 cc/sec for 22Fr, and − 0.055 to 1.243 cc/sec for 24Fr. In case of maximum … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...

Citation Types

0
0
0

Year Published

2024
2024
2024
2024

Publication Types

Select...
1

Relationship

0
1

Authors

Journals

citations
Cited by 1 publication
references
References 17 publications
0
0
0
Order By: Relevance