“…The initial generations of ICDs did not record and save electrograms (EGMs), leading to a reduced appreciation for the frequency and impact of inappropriate shocks. With the advent and then dominance of primary prevention indications, avoidable shocks assumed a relatively larger proportion of total therapy [78][79][80][81][82][83]. Gradually, publications have increased awareness of the frequency and the diverse range of adverse outcomes associated with avoidable ICD therapy, and have demonstrated that avoidable ICD shocks can be reduced by evidence-based programming of the detection rate, detection duration, antitachycardia pacing (ATP), algorithms that discriminate supraventricular tachycardia (SVT) from VT, and specific programming to minimize the sensing of noise [81][82][83][84][85][86][87][88][89][90][91][92].…”