Abstract:This research aimed to evaluate the effects of the X-ray current tube on the uniformity value of CT image noise. This research used the CT images of Siemens head water phantom. The phantom was scanned with a Siemens Somatom Scope CT scanner for various current tubes (i.e., 180 mAs, 200 mAs, 220 mAs, 240 mAs, and 260 mAs) on tube voltage 110 kV, slice thickness 3 mm, and FOV 240 mm. The uniformity values in CT image noise were analyzed using the standard deviation value (SD). The SD was measured by the ROI proc… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.