“…It can be observed that for PCT-BR and PCT-BK, the between-day component of imprecision was negligible, whereas for the PCT-D, especially at lower concentrations, the between-day component was relevant. The observed LoD and the functional sensitivity for PCT-D were much higher than those declared by Diazyme (0.051 and 0.15 μg/L) [19], but in line with those published by Dipalo et al [20], whereas for PCT-BR and PCT-BK, the values were slightly better than those declared [17,18]. The reproducibility performances of PCT-BR at very low PCT concentrations are clearly superior to those of the other two systems (Figure 1).…”