2016
DOI: 10.1016/j.rpor.2015.10.004
|View full text |Cite
|
Sign up to set email alerts
|

Analyzing the performance of ArcCHECK diode array detector for VMAT plan

Abstract: The VMAT patient specific QA with an ion chamber and ArcCHECK phantom are consistent with the TPS calculated dose. Statistically good agreement was observed between ArcCHECK measured and TPS calculated. Hence, it can be used for routine VMAT QA.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
16
0
1

Year Published

2016
2016
2022
2022

Publication Types

Select...
10

Relationship

0
10

Authors

Journals

citations
Cited by 24 publications
(17 citation statements)
references
References 30 publications
0
16
0
1
Order By: Relevance
“…Pretreatment patient-specific delivery quality assurance (DQA) is essential for the verification of accurate dose delivery in advanced radiation therapy techniques, such as intensity-modulated radiation therapy (IMRT), volumetric modulated arc therapy, stereotactic body radiation therapy, and helical tomotherapy (HT). 1 - 4 …”
Section: Introductionmentioning
confidence: 99%
“…Pretreatment patient-specific delivery quality assurance (DQA) is essential for the verification of accurate dose delivery in advanced radiation therapy techniques, such as intensity-modulated radiation therapy (IMRT), volumetric modulated arc therapy, stereotactic body radiation therapy, and helical tomotherapy (HT). 1 - 4 …”
Section: Introductionmentioning
confidence: 99%
“…All 20 plans were measured using a Sun Nuclear ArcCHECK, a 21×21×15 cm3 cylindrical dose measurement tool which consists of 1,386 0.019 mm 2 diode detectors each spaced 1.0 cm apart (13) . Prior studies have validated the ArcCHECK as an IMRT and VMAT quality assurance (QA) measuring tool (e.g., Nelms et al (14) ).…”
Section: Methodsmentioning
confidence: 99%
“…It is a cylindrical water-equivalent phantom with a three-dimensional array of 1386 diode detectors, arranged in a spiral pattern, with 10 mm sensor spacing. Dose measurements from each sensor were updated every 50ms [29]. FIGURE 4 shows ArcCHECK on the motion platform.…”
Section: ) 3d Dynamic Dose Verificationmentioning
confidence: 99%