Angular dependency correction of 2D planar detector I’mRT MatriXX an offline dosimetry system used for IMRT pre-treatment verification

Abstract: I'mRT MatriXX is one of the common offline dosimetry tools used for the pre-treatment verification quality assurance (QA) of intensity modulated radiation therapy (IMRT) treatment plans in radiotherapy. It has 1020 pixel ionisation chambers (PIC) arranged in 32×32 grid in a plain. In this study the angular response of the I'mRT MatriXX was measured using treatment planning system (TPS) calculated dose and I'mRT MatriXX measured dose. The angular response correction factor (CF) of the I'mRT MatriXX was found … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

1
10
0

Year Published

2021
2021
2023
2023

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(11 citation statements)
references
References 10 publications
1
10
0
Order By: Relevance
“…The asymmetry about 90 • of the Δd(θ) agrees well with the asymmetric about 90 • of the CF profiles. The reproduce of the CF profile in Figure 2b using the DEMP model showed analogous shape to those of previous works, 7,8,13 verifying the efficacy of DEMP model.The fact that neither the depth offset curve of the EMP nor the CF curve is much influenced by the simulated set up error of 1 mm in Figure 2a,b demonstrated the robustness of the DEMP model to the set up uncertainties. Additionally, the derivatives of CF curve in Figure 2c revealed that the sensitivity of CF to the gantry angle is largest in the gantry angle range of (100 • , 120 • ).…”
Section: Discussionsupporting
confidence: 82%
See 4 more Smart Citations
“…The asymmetry about 90 • of the Δd(θ) agrees well with the asymmetric about 90 • of the CF profiles. The reproduce of the CF profile in Figure 2b using the DEMP model showed analogous shape to those of previous works, 7,8,13 verifying the efficacy of DEMP model.The fact that neither the depth offset curve of the EMP nor the CF curve is much influenced by the simulated set up error of 1 mm in Figure 2a,b demonstrated the robustness of the DEMP model to the set up uncertainties. Additionally, the derivatives of CF curve in Figure 2c revealed that the sensitivity of CF to the gantry angle is largest in the gantry angle range of (100 • , 120 • ).…”
Section: Discussionsupporting
confidence: 82%
“…Previous methods accounting for the angular dependency of MatriXX response mostly were based on some sort of CFs. 7 , 8 , 13 , 14 It should be noted that the DEMP model in this study is to calculate the angularly corrected composite planar dose, which is to be compared with the MatriXX response, using the TPS‐calculated dose distribution from each beam; while the conventional CF methods is to correct the MatriXX response from each beam and then synthesize them to match the TPS‐calculated dose on the measurement plane. The distinction between the two methods could be summarized as the conventional CF method corrects the MatriXX response, while the DEMP‐based approach corrects the TPS‐calculated planar dose.…”
Section: Discussionmentioning
confidence: 99%
See 3 more Smart Citations