Anterior variable-angle locked plating versus tension band wiring of simple and complex patella fractures – a biomechanical investigation

Abstract: Background The aim of this study was to investigate the biomechanical performance of novel anterior variable-angle locking plates versus tension band wiring used for fixation of simple and complex patella fractures. Methods Sixteen pairs of human cadaveric knees were used to simulate two-part simple transverse AO/OTA 34-C1 and five-part complex AO/OTA 34-C3 patella fractures. The complex fracture pattern was characterized with a medial and a latera… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
4
0

Year Published

2023
2023
2024
2024

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 10 publications
(4 citation statements)
references
References 49 publications
0
4
0
Order By: Relevance
“…Fixed-angle plates, in particular, have been proven to provide better biomechanical support and reduce the widening of fractures during movement compared to using cannulated screws with the VA LNP technique [ 29 ]. Recent designs of VA LNP are believed to offer enhanced adjustability and stability [ 13 , 14 , 30 ]. Stoffel et al studied the effectiveness of using TBW versus using VA LNP in cadaver tests, and the VA LNP approach showed significantly less displacement after cycling compared to the tension band wiring method [ 14 ].…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…Fixed-angle plates, in particular, have been proven to provide better biomechanical support and reduce the widening of fractures during movement compared to using cannulated screws with the VA LNP technique [ 29 ]. Recent designs of VA LNP are believed to offer enhanced adjustability and stability [ 13 , 14 , 30 ]. Stoffel et al studied the effectiveness of using TBW versus using VA LNP in cadaver tests, and the VA LNP approach showed significantly less displacement after cycling compared to the tension band wiring method [ 14 ].…”
Section: Discussionmentioning
confidence: 99%
“…Recent designs of VA LNP are believed to offer enhanced adjustability and stability [ 13 , 14 , 30 ]. Stoffel et al studied the effectiveness of using TBW versus using VA LNP in cadaver tests, and the VA LNP approach showed significantly less displacement after cycling compared to the tension band wiring method [ 14 ]. Similar findings were reported by Kfuri et al [ 30 ] However, there is a lack of data on the effectiveness of these VA LNPs for simple transverse patella fractures.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…are the preferred surgical treatments for inferior pole fractures of patella. [8,9] Resection of the inferior pole of patella is rarely applied at present because it changes the biomechanical characteristics of the patella, causes weakened quadriceps muscle strength and patellofemoral arthritis and affects the function of the knee joint. In contrast, Kirschner wire tension band internal fixation is widely applied.…”