Approximation of a melting plateau of large area HTFP cells used for spectral irradiance realization

Abstract: This paper adopted an approximation of a melting plateau to solve the problem that temperature data cannot be monitored continuously when measuring the spectral irradiance of a large area tungsten carbide–carbon high-temperature fixed-point blackbody at each measured wavelength. Tests with fully measured curves showed that the method has a rather small deviation from the measured data of 0.017 K maximum, which corresponds to the spectral irradiance deviation of 0.005% at 500 … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1

Citation Types

0
3
0

Year Published

2022
2022
2022
2022

Publication Types

Select...
1

Relationship

1
0

Authors

Journals

citations
Cited by 1 publication
(3 citation statements)
references
References 28 publications
0
3
0
Order By: Relevance
“…When the second melting plateau appeared, the optical platform was moved quickly to the position of the spectral comparator to measure the signal of the HTFP at several wavelengths. When the measurement was completed, the optical platform was quickly moved back to the position of the pyrometer to monitor the end of the melting curve [20,22]. Because of limited number of spectral point could be measure by a comparator during one melting plateau, a few melt/freeze cycles were done to cover the whole spectral range from 250 nm to 2500 nm.…”
Section: Measurement Proceduresmentioning
confidence: 99%
See 2 more Smart Citations
“…When the second melting plateau appeared, the optical platform was moved quickly to the position of the spectral comparator to measure the signal of the HTFP at several wavelengths. When the measurement was completed, the optical platform was quickly moved back to the position of the pyrometer to monitor the end of the melting curve [20,22]. Because of limited number of spectral point could be measure by a comparator during one melting plateau, a few melt/freeze cycles were done to cover the whole spectral range from 250 nm to 2500 nm.…”
Section: Measurement Proceduresmentioning
confidence: 99%
“…The Akima fitting method was used to restore the missing part of the melting plateau curve [22]. Figure 6 shows the fitting result: the measured data at the beginning and the end of the melting plateau, and the central recovered data by using the Akima fitting method.…”
Section: Fitting and Correcting Of Melting Plateaumentioning
confidence: 99%
See 1 more Smart Citation