Both analyzers provided precise results of pH, PaCO , PaO and HCT, meeting CV% quality requirement values. In both species, Deming regression results were acceptable and correlation values were above 0.93 for arterial pH and blood gases, but lower for sodium and chloride. Bland-Altman plots demonstrated varying degrees of bias, but good agreement between the 2 analyzers was seen when arterial blood gases and electrolytes were measured, except for PaCO and Cl CONCLUSION: The EPOC analyzer provides consistent, reliable results for canine arterial blood gas values and for equine arterial blood gas and electrolyte values. Cl results could be acceptable with the application of a correction factor, but the PaCO results were more variable.