A portable flow analysis instrument is described for in situ
photometric measurements. This system is based on light-emitting
diodes (LEDs) and a photodiode detector, coupled to a multipumping
flow system. The whole equipment presents dimensions of 25 cm × 22 cm ×
10 cm, weighs circa 3 kg,
and costs 650 €. System performance was evaluated for
different chemistries without changing hardware configuration for
determinations of (i) Fe3+ with SCN-, (ii)
iodometric nitrite determination, (iii) phenol with sodium
nitroprusside, and (iv) 1-naphthol-N-methylcarbamate
(carbaryl) with p-aminophenol. The detection limits were
estimated as 22, 60, 25, and 60 ng mL -1 for iron,
nitrite, phenol, and carbaryl at the 99.7% confidence level with
RSD of 2.3, 1.0, 1.8, and 0.8%, respectively. Reagent and waste
volumes were lower than those obtained by flow systems with
continuous reagent addition. Sampling rates of 100, 110, 65, and
72 determinations per hour were achieved for iron, nitrite,
phenol, and carbaryl determinations