Background and objectivesBioelectrical impedance analysis (BIA) devices can help assess volume overload in patients receiving maintenance peritoneal dialysis. However, the effects of BIA on the short-term hard end points of peritoneal dialysis lack consistency. This study aimed to test whether BIA-guided fluid management could improve short-term outcomes in patients on peritoneal dialysis.Design, setting, participants, & measurementsA single-center, open-labeled, randomized, controlled trial was conducted. Patients on prevalent peritoneal dialysis with volume overload were recruited from July 1, 2013 to March 30, 2014 and followed for 1 year in the initial protocol. All participants with volume overload were 1:1 randomized to the BIA-guided arm (BIA and traditional clinical methods) and control arm (only traditional clinical methods). The primary end point was all-cause mortality and secondary end points were cardiovascular disease mortality and technique survival.ResultsA total of 240 patients (mean age, 49 years; men, 51%; diabetic, 21%, 120 per group) were enrolled. After 1-year follow-up, 11(5%) patients died (three in BIA versus eight in control) and 21 patients were permanently transferred to hemodialysis (eight in BIA versus 13 in control). The rate of extracellular water/total body water decline in the BIA group was significantly higher than that in the control group. The 1-year patient survival rates were 96% and 92% in BIA and control groups, respectively. No significant statistical differences were found between patients randomized to the BIA-guided or control arm in terms of patient survival, cardiovascular disease mortality, and technique survival (P>0.05).ConclusionsAlthough BIA-guided fluid management improved the fluid overload status better than the traditional clinical method, no significant effect was found on 1-year patient survival and technique survival in patients on peritoneal dialysis.