Design: cadaveric spine nucleus replacement study. Objective:
determining
Bionate 80A nucleus replacement biomechanics in cadaveric spines.
Methods: in cold preserved spines, with ligaments and discs intact,
and no muscles, L3-L4, L4-L5, and L5-S1 nucleus implantation was done.
Differences between customized and overdimensioned implants were compared.
Flexion, extension, lateral bending, and torsion were measured in
the intact spine, nucleotomy, and nucleus implantation specimens.
Increasing load or bending moment was applied four times at 2, 4,
6, and 8 Nm, twice in increasing mode and twice in decreasing mode.
Spine motion was recorded using stereophotogrammetry. Expulsion tests:
cyclic compression of 50–550 N for 50,000 cycles, increasing
the load until there was extreme flexion, implant extrusion, or anatomical
structure collapse. Subsidence tests were done by increasing the compression
to 6000 N load. Results: nucleotomy increased the disc mobility, which
remained unchanged for the adjacent upper level but increased for
the lower adjacent one, particularly in lateral bending and torsion.
Nucleus implantation, compared to nucleotomy, reduced disc mobility
except in flexion-extension and torsion, but intact mobility was no
longer recovered, with no effect on upper or lower adjacent segments.
The overdimensioned implant, compared to the customized implant, provided
equal or sometimes higher mobility. Lamina, facet joint, and annulus
removal during nucleotomy caused more damaged than that restored by
nucleus implantation. No implant extrusion was observed under compression
loads of 925–1068 N as anatomical structures collapsed before.
No subsidence or vertebral body fractures were observed under compression
loads of 6697.8–6812.3 N. Conclusions: nucleotomized disc and
L1-S1 mobility increased moderately after cadaveric
spine nucleus implantation compared to the intact status, partly due
to operative anatomical damage. Our implant had shallow expulsion
and subsidence risks.