Biomechanical Stress Analysis of Platform Switch Implants of Varying Diameters on Different Densities of Bone

Abstract: Purpose. The purpose of this study was to evaluate and compare the strain developed in D2 and D3 types of bones on vertical loading by platform switch implants of different diameters. Materials and Methods. Implants of diameters 3.25 mm, 4.2 mm, and 5.0 mm and of length 11.5 mm were taken and placed each on D2 and D3 bone models. Strain gauges were attached on the buccal and the lingual sides on each of these samples, and a vertical load of 190N was placed on the samples. The strain was recorded using a data l… Show more