There is fair evidence to recommend against the use of high offset femoral heads, larger diameter femoral heads, and to pay particular consideration to fretting corrosion's progression with time and risk with heavier or more active patients. Particular to metal-on-metal hip prostheses, there is fair evidence to recommend positioning the acetabular component to minimise edge loading. Particular to metal-on-polyethylene hip prostheses, there is fair evidence to recommend the use of ceramic femoral heads, against use of cast cobalt alloy femoral heads, and against use of low flexural rigidity femoral stems. Evidence related to taper connection design is largely conflicting or inconclusive. Head-neck taper connection fretting corrosion is a multifactorial problem. Strict adherence to the guidelines presented herein does not eliminate the risk. Prosthesis selection is critical, and well-controlled studies to identify each design parameter's relative contribution to head-neck taper connection fretting corrosion are required.