2015
DOI: 10.1115/1.4031891
|View full text |Cite
|
Sign up to set email alerts
|

Bond Strength of Thermally Fused Vascular Tissue Varies With Apposition Force

Abstract: Surgical tissue fusion devices ligate blood vessels using thermal energy and coaptation pressure, while the molecular mechanisms underlying tissue fusion remain unclear. This study characterizes the influence of apposition force during fusion on bond strength, tissue temperature, and seal morphology. Porcine splenic arteries were thermally fused at varying apposition forces (10-500 N). Maximum bond strengths were attained at 40 N of apposition force. Bonds formed between 10 and 50 N contained laminated medial … Show more