The aim of this study was to devise an algorithm that would predict flow cytometry crossmatch (FCXM) results using single‐antigen bead (SAB) mean fluorescent intensity (MFI) levels using samples received through the National External Quality Assurance Scheme (NEQAS) 2B external proficiency testing scheme between 2019 and 2023. A total of 159 serum samples were retrospectively screened using LABScreen Single Antigen Class I and II (SAB), and 40 peripheral blood samples were human leucocyte antigen (HLA) typed with LABType SSO. Donor‐specific antibodies were identified for each cell–serum combination tested, and cumulative MFI values were calculated for each test before correlating the screening result with the consensus crossmatch results for this scheme. HLA Class I MFIs were combined to predict the T cell crossmatch. For the B cell crossmatch prediction, two options were considered: (i) HLA Class II MFI values alone and (ii) HLA Class I + Class II MFIs. Receiver operating characteristic analysis was carried out to identify the combined MFI threshold that predicted NEQAS consensus results with the greatest sensitivity and specificity. HLA Class I combined MFI >5000 predicted T cell crossmatch results with 96% sensitivity, 100% specificity, 100% positive predictive value (PPV) and 92% negative predictive value (NPV). For B cell results, HLA Class I + Class II combined MFIs >11,000 gave the best model, showing 97% sensitivity, 82% specificity, 96% PPV and 85% NPV. However, for samples with only HLA Class II sensitization, combined MFIs >13,000 improved the B cell crossmatch predictions: 92% sensitivity, 95% specificity, 96% PPV and 91% NPV. Using this model, combined MFI can be used to predict the immunological risk posed by donor‐specific antibodies when it is not possible to carry out an FCXM.