2017
DOI: 10.1016/j.radmeas.2017.06.011
|View full text |Cite
|
Sign up to set email alerts
|

Calculation of thermal quenching parameters in BeO ceramics using solely TL measurements

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
4
0

Year Published

2017
2017
2024
2024

Publication Types

Select...
7

Relationship

1
6

Authors

Journals

citations
Cited by 21 publications
(5 citation statements)
references
References 31 publications
1
4
0
Order By: Relevance
“…According to Kitis and Pagonis [ 45 ], in such cases, the large number of re-trappings indicates TL glow curves that are also dominated by the first order of kinetics. The activation energies are slightly larger than 1 eV, in good agreement with the results of the activation energies reported by our group for both TL peaks 2 and 3 [ 4 , 5 , 14 , 15 , 40 ]. Moreover, following TA-OSL, there is no change in the TL glow curve shape of BeO; such a change was quite obvious in both cases of Al 2 O 3 :C and NaCl.…”
Section: Discussionsupporting
confidence: 91%
See 1 more Smart Citation
“…According to Kitis and Pagonis [ 45 ], in such cases, the large number of re-trappings indicates TL glow curves that are also dominated by the first order of kinetics. The activation energies are slightly larger than 1 eV, in good agreement with the results of the activation energies reported by our group for both TL peaks 2 and 3 [ 4 , 5 , 14 , 15 , 40 ]. Moreover, following TA-OSL, there is no change in the TL glow curve shape of BeO; such a change was quite obvious in both cases of Al 2 O 3 :C and NaCl.…”
Section: Discussionsupporting
confidence: 91%
“…For the case of isothermal TA-OSL, as the entire measurement takes place at a steady temperature T st , the corrected unquenched integrated TA-OSL intensity I uq ( T st ) can be calculated using the following formula: where I q ( T st ) and I uq ( T st ) are the integrated TA-OSL intensities which are quenched and unquenched, respectively, while η ( T st ) corresponds to a single value of the thermal quenching efficiency at the specific measurement temperature. In this study, reconstruction was applied according to Equation (11), using the values of thermal quenching parameters W and C obtained by Aşlar et al [ 40 ] for the Hoya U340 filter; the specific values used were W = 0.62 eV and C = 1.3 × 10 6 , namely, the values that were calculated for TL peak 2. These values were used because the Hoya U-340 filter was used for all measurements.…”
Section: Methodsmentioning
confidence: 99%
“…The dimensions and weight of the dosimeter are 4 mm × 4 mm × 1 mm and 32 mg, respectively. This dosimeter was already characterized in terms of TL and OSL kinetics and dose response [1–3]. Indeed, BeO is one of the most promising OSL materials for radiation monitoring and dosimetery.…”
Section: Methodsmentioning
confidence: 99%
“…BeO dosimeters exhibit high sensitivity to ionizing radiation and are reusable, allowing for multiple readouts. OSL and TL kinetics and dose response of the studied BeO dosimeter were extensively studied showing two experimental main peaks recorded at a temperature interval of 200–370°C depending on heating rate and material preparation [1–3]. Regarding the TL response as a function of dose, recent studies show that the first peak obeys to second‐order kinetics and exhibits a quadratic dependence to dose while the second peak is of first‐order kinetics with a linear behaviour with respect to dose [4, 5].…”
Section: Introductionmentioning
confidence: 99%
“…Therefore, thermal quenching studies are of great importance not only for candidate dosimetric materials but also for the dosimeters in luminescence dosimetry research. The existence of a thermal quenching effect can be identified by monitoring the TL glow curves obtained at different heating rates under the same amount of radiation exposure [4]. Thermal quenching can also be understood with the change in the OSL signal obtained at different reading temperatures and luminescence lifetime measurements [5][6][7][8][9].…”
Section: Introductionmentioning
confidence: 99%