Calibration and analysis of a multimodal micro-CT and structured light imaging system for the evaluation of excised breast tissue

Abstract: A multimodal micro-computed tomography (CT) and multi-spectral structured light imaging (SLI) system is introduced and systematically analyzed to test its feasibility to aid in margin delineation during breast conserving surgery (BCS). Phantom analysis of the micro-CT yielded a signal-to-noise ratio (SNR) of 34, a contrast of 1.64, and a minimum detectable resolution of 240 µm for a 1.2 min scan. The SLI system, spanning wavelengths 490 nm to 800 nm and spatial frequencies up to 1.37 mm−1, was evaluated with a… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
20
0

Year Published

2018
2018
2023
2023

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 24 publications
(22 citation statements)
references
References 48 publications
0
20
0
Order By: Relevance
“…49,50 Micro-CT has also been combined with other types of imaging systems such as spatial frequency domain imaging (SFDI) to create multidimensional data that may offer accurate specimen margin assessments. 51 A recent study of fresh lumpectomy specimens was able to identify the correct margin status on 25 out of 29 patients that were able to be analyzed. There were four false negative results which lead to a sensitivity of 56% and a specificity of 100%.…”
Section: Micro-ct (3-d Volumes)mentioning
confidence: 99%
“…49,50 Micro-CT has also been combined with other types of imaging systems such as spatial frequency domain imaging (SFDI) to create multidimensional data that may offer accurate specimen margin assessments. 51 A recent study of fresh lumpectomy specimens was able to identify the correct margin status on 25 out of 29 patients that were able to be analyzed. There were four false negative results which lead to a sensitivity of 56% and a specificity of 100%.…”
Section: Micro-ct (3-d Volumes)mentioning
confidence: 99%
“…While this sort of behavior is not necessarily fundamentally flawed (this is the expected behavior of the algorithm), for in vivo imaging this implies that contrast will fluctuate as different sorts of spectra are presented to the camera in various proportions. For example, in the case of margin delineation, tumors may be accentuated if surrounded by adipose tissue, or darkened if surrounded by normal or connective tissue (adipose and normal tissue can present lower and higher scattering coefficients and average reflectance than malignant tissue, respectively) [15,34]. These inconsistencies may be problematic in a clinical setting, explaining the motivation behind this desired improvement, and providing an insight on how these techniques cannot be exploited for specific applications unless the dependency with the SVD is avoided.…”
Section: Susceptibility Of Current Methods To Changes In the Environmentmentioning
confidence: 99%
“…Its performance was quantified experimentally, as described in detail previously. 12 The SLI system demonstrated a maximum spatial resolution of 3.78 cycles∕mm at the highest illumination spatial frequency (f x ) of 1.37 mm −1 and at the shortest source wavelength (λ) of 490 nm. Optical scattering properties were recovered accurately with ratios of μ 0 s ∕μ a as low as 0.8.…”
Section: Instrumentationmentioning
confidence: 99%
“…The system imaged tissue specimens with four illumination spatial frequencies (f x ¼ 0.00, 0.15, 0.60, and 1.37 mm −1 ) at four source wavelengths (λ ¼ 490, 550, 600, 700 nm). 12 It acquired a lateral cone beam μCT scan. The analysis presented here is focused solely on the SLI data.…”
Section: Instrumentationmentioning
confidence: 99%
See 1 more Smart Citation