Purpose
Different non-invasive and invasive imaging modalities are used to determine carotid artery stenosis severity that remains a principal parameter in clinical decision-making. We compared stenosis degree obtained with different modalities against vascular imaging gold standard, intravascular ultrasound, IVUS.
Methods
300 consecutive patients (age 47–83 years, 192 men, 64% asymptomatic) with carotid artery stenosis of “ ≥ 50%” referred for potential revascularization received as per study protocol (i) duplex ultrasound (DUS), (ii) computed tomography angiography (CTA), (iii) intraarterial quantitative angiography (iQA) and (iv) and (iv) IVUS. Correlation of measurements with IVUS (r), proportion of those concordant (within 10%) and proportion of under/overestimated were calculated along with recipient-operating-characteristics (ROC).
Results
For IVUS area stenosis (AS) and IVUS minimal lumen area (MLA), there was only a moderate correlation with DUS velocities (peak-systolic, PSV; end-diastolic, EDV; r values of 0.42–0.51, p < 0.001 for all). CTA systematically underestimated both reference area and MLA (80.4% and 92.3% cases) but CTA error was lesser for AS (proportion concordant-57.4%; CTA under/overestimation-12.5%/30.1%). iQA diameter stenosis (DS) was found concordant with IVUS in 41.1% measurements (iQA under/overestimation 7.9%/51.0%). By univariate model, PSV (ROC area-under-the-curve, AUC, 0.77, cutoff 2.6 m/s), EDV (AUC 0.72, cutoff 0.71 m/s) and CTA-DS (AUC 0.83, cutoff 59.6%) were predictors of ≥ 50% DS by IVUS (p < 0.001 for all). Best predictor, however, of ≥ 50% DS by IVUS was stenosis severity evaluation by automated contrast column density measurement on iQA (AUC 0.87, cutoff 68%, p < 0.001). Regarding non-invasive techniques, CTA was the only independent diagnostic modality against IVUS on multivariate model (p = 0.008).
Conclusion
IVUS validation shows significant imaging modality-dependent variations in carotid stenosis severity determination.