The EHRA Book of Pacemaker, ICD and CRT Troubleshooting Vol. 2 2022
DOI: 10.1093/med/9780192844170.003.0057
|View full text |Cite
|
Sign up to set email alerts
|

Case 57

Abstract: A patient complained of persistent heart failure symptoms 6 months after CRT implantation. Device follow-up showed excellent capture thresholds with 99% VP and no sustained arrhythmias. The paced AVI had been programmed at the default value of 130 ms.