Instrument breakage during orthopaedic procedures is relatively uncommon, but it remains a challenging problem to orthopaedic surgeons. We report a case of a broken guidewire during intramedullary nail fixation of a proximal femur fracture with subtrochanteric extension. We also report a simple yet effective technique for removal of a broken guidewire from the medulla of the proximal femur without causing much iatrogenic trauma.