Characterizing current noise of commercial constant-current sources by using an optically pumped rubidium atomic magnetometer
Ni Zhao,
Lulu Zhang,
Yongbiao Yang
et al.
Abstract:This paper introduces a method for characterizing the current noise of commercial constant-current sources (CCSs) using a free-induction-decay (FID) type optically pumped rubidium atomic magnetometer driven by a radio frequency magnetic field. We convert the sensitivity of the atomic magnetometer into the current noise of CCS by calibrating the coil constant. At the same time, the current noise characteristics of six typical commercial low-noise CCSs are compared. The current noise level of the Keysight model … Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.