Abstract:A flow injection analysis (FIA) system having a chlormequat selective electrode is proposed. Several electrodes with poly(vinyl chloride) based membranes were constructed for this purpose. Comparative character-ization suggested the use of membrane with chlormequat tetraphenylborate and dibutylphthalate. On a single-line FIA set-up, operating with 1 x 10 -2 mol L -1 ionic strength and 6.3 pH, calibration curves presented slopes of 53.6 ± 0.4 mV decade -1 within 5.0 x 10 -6 and 1.0 x 10 -3 mol L -1, and squared… Show more
O PDF relativo ao artigo que solicita, não se encontra disponível em Acesso Aberto.
Motivos: O editor não permite o depósito e disponibilização em acesso aberto do PDF que solicita. Para consultar o documento deve aceder ao endereço do editor.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.