2008
DOI: 10.1007/s00264-008-0678-9
|View full text |Cite
|
Sign up to set email alerts
|

Clinical and biological assessment of cemented titanium femoral stems: an 11-year experience

Abstract: This study prospectively assessed the outcome of 134 cemented titanium stems and serum ion levels. The stems were polished (0.1 μm Ra) with circular cross section. At the end point, only one stem revision was performed for aseptic loosening, and two were planned due to subsidence greater than 5 mm. Non-progressive radiolucencies in zones 1 and 7 were observed in 16 hips at the cement-bone interface without osteolysis. Median serum titanium concentrations were below the detection limit (30 nmol/l) except in pat… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
10
0
2

Year Published

2009
2009
2018
2018

Publication Types

Select...
6
2

Relationship

1
7

Authors

Journals

citations
Cited by 16 publications
(13 citation statements)
references
References 27 publications
1
10
0
2
Order By: Relevance
“…We did not find any evidence to support these hypotheses, however. First of all, the performance of the stem was excellent, with 99% survivorship at the last review; these results are in accordance with the results of previously published studies ( Nich et al 2003 , Boyer et al 2008 ). The only elevated serum titanium levels were the result of stem loosening due to poor cementing technique or to neck impingement (in the 2 cases of dislocation).…”
Section: Discussionsupporting
confidence: 91%
“…We did not find any evidence to support these hypotheses, however. First of all, the performance of the stem was excellent, with 99% survivorship at the last review; these results are in accordance with the results of previously published studies ( Nich et al 2003 , Boyer et al 2008 ). The only elevated serum titanium levels were the result of stem loosening due to poor cementing technique or to neck impingement (in the 2 cases of dislocation).…”
Section: Discussionsupporting
confidence: 91%
“…In short, it is advantageous to have fit-for-purpose, accurate analytical techniques for Ti in whole blood to allow monitoring of exposures. As can be seen in Table 1, most commonly, inductively coupled plasma mass spectrometry (ICP-MS) has been used for such measurements 1,14, 17 -64 although ICP optical emission spectrometry (ICP-OES) 23,26,62,63,[65][66][67][68][69] and non-ICP techniques like graphite furnace atomic absorption spectrometry (GFAAS) have also been applied 24, 35, 69 -73 . In such analyses whole blood may be diluted and analysed directly or digested before analysis albeit, as noted above, only HF and hot sulphuric acid are likely to provide complete digestion of TiO 2 12 .…”
Section: Introductionmentioning
confidence: 99%
“…Historically, the performance of cemented titanium alloy femoral stems has been mixed. They are generally considered to perform unsatisfactorily; however, some designs have survivorship ranging from 97.7% at 9 years to 95.4% at 13 years [3,4,11]. In 2008, surgeons identified a change in the performance of the Titan 1 stem, and the NAR later reported a drop in survivorship due to aseptic loosening from about 2000 [5].…”
Section: Discussionmentioning
confidence: 99%