Abstract:Objective: Evaluate the technical performance of the pre-analytical hemolysis-icterus-lipemia (HIL) check module on the ACL-TOP-750. Methods: 8433 routine coagulation samples were evaluated for HIL, the presence of clotting and low sample volume by both visual inspection and the pre-analytical HIL check module on the ACL-TOP-750. Results: 7726 samples were in agreement with both methods and 707 were not consistent. 356 samples with low volume were identified by visual inspection and 920 by the instrument (2.7 … Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.