Clinical feasibility of a novel test setup for objective measurements using the VIBRANT SOUNDBRIDGE

Abstract: Objectives The VIBRANT SOUNDBRIDGE is a widely used active middle ear implant to treat hearing loss. The floating mass transducer is surgically coupled to the ossicles, the round or oval window. A reliable method to monitor the coupling efficiency intraoperatively is highly desired. Research groups have developed several methods, but limitations remain. This study aims to evaluate the clinical feasibility of a new research setup for auditory brainstem response measurement to evaluate the coupling efficiency. M… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

2
3
0

Year Published

2023
2023
2023
2023

Publication Types

Select...
4

Relationship

0
4

Authors

Journals

citations
Cited by 4 publications
(5 citation statements)
references
References 16 publications
2
3
0
Order By: Relevance
“…It should be noted, however, that direct comparisons between our study and that of Fröhlich et al should be made with caution because Fröhlich et al used a preprogrammed sound processor with an attached insert earphone for each implantation, resulting in different intraoperative ABR measurements. A recent study conducted by Sprinzl et al presented findings similar to our study in terms of test design, measurements, cohort, and data analysis with 14 AMEI implantations ( 16 ). Our results showed an intraoperative ABR threshold almost identical to theirs.…”
Section: Discussionsupporting
confidence: 89%
See 2 more Smart Citations
“…It should be noted, however, that direct comparisons between our study and that of Fröhlich et al should be made with caution because Fröhlich et al used a preprogrammed sound processor with an attached insert earphone for each implantation, resulting in different intraoperative ABR measurements. A recent study conducted by Sprinzl et al presented findings similar to our study in terms of test design, measurements, cohort, and data analysis with 14 AMEI implantations ( 16 ). Our results showed an intraoperative ABR threshold almost identical to theirs.…”
Section: Discussionsupporting
confidence: 89%
“…Fröhlich et al conducted a study of 18 patients with similar demographic and audiological characteristics to our cohort (17). They investigated the frequency-specific coupling efficiency and (16). Our results showed an intraoperative ABR threshold almost identical to theirs.…”
Section: Abrsupporting
confidence: 80%
See 1 more Smart Citation
“…The intraoperative ABR measurements were performed using an Interacoustics Eclipse device as previously described [ 20 , 21 ]. The system allowed for the generation of the stimulus and the recording of the brainstem potentials.…”
Section: Methodsmentioning
confidence: 99%
“…Sprinzl et al (27) investigated coupling efficiency in vibroplasty using intraoperative ABR measurements in 13 patients. The mean difference between preoperative bone conduction thresholds and intraoperative ABR thresholds was 7 dB (SD, ±10 dB), which is slightly higher but still within the 10-dB range.…”
Section: Discussionmentioning