The research presented here describes the development of two analytical methods for use in the certification of a ginsenoside calibration solution Standard Reference Material (SRM) 3389 consisting of seven ginsenosides: Rg1, Re, Rf, Rb1, Rc, Rb2, and Rd. The new methods utilized the liquid chromatographic (LC) separation of ginsenoside mixtures with absorbance detection (UV) and mass spectrometry (MS). Ginsenosides Rb3, Rg2, Rg3, Rh1, and Rh2 were evaluated for use as internal standards for LC/MS measurements. The 12 ginsenosides were baseline resolved by gradient elution LC/UV, with an initial mobile phase composition of 22 % acetonitrile and 78 % water, flow rate of 0.7 mL/min, and column temperature of 25 °C. The work presented here includes a detailed investigation into the optimization of the chromatographic conditions to minimize measurement biases that result from unresolved constituents. Temperature and mobile phase composition are known to play a significant role in column selectivity; however, flow rate is expected to influence primarily the separation efficiency and detection sensitivity. In the current study, column selectivity changed with changes in flow rate and the relative retention of ginsenoside Rg2 and Rh1 changed as the flow rate increased from 0.6 mL/min to 1.0 mL/min.