2021
DOI: 10.1159/000518357
|View full text |Cite
|
Sign up to set email alerts
|

Comparison of Anorectal Function as Measured with High-Resolution and High-Definition Anorectal Manometry

Abstract: <b><i>Introduction:</i></b> Anorectal manometry (ARM) provides comprehensive assessment of pressure activity in the rectum and anal canal. Absolute pressure values might depend on the catheter used. <b><i>Objective:</i></b> Our aim was to compare the results obtained by different anorectal catheters in children with functional anorectal disorders. <b><i>Methods:</i></b> Children diagnosed with functional defecation disorders based on the R… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
4
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
5

Relationship

1
4

Authors

Journals

citations
Cited by 5 publications
(4 citation statements)
references
References 46 publications
0
4
0
Order By: Relevance
“…Two recent prospective studies reported a higher anal resting pressure recorded and conflictive RAIR findings when 3D high-resolution ARM (HRARM) is used compared to 2D HRARM consecutively on the same patients. 63 However, Keshtgar et al in another recent study reported no significant difference in values of anal resting pressure and volume threshold to elicit RAIR obtained with conventional water perfusion or 3D high-resolution ARM. 64 ARM guidelines for performance and interpretation were revised and published by the American Neurogastroenterology and Motility Society in conjunction with The North American Society of Paediatric Gastroenterology, Hepatology and Nutrition 65 and recently by the British Society of Paediatric Gastroenterology, Hepatology and Nutrition (BSPGHAN).…”
Section: Study Performancementioning
confidence: 93%
See 1 more Smart Citation
“…Two recent prospective studies reported a higher anal resting pressure recorded and conflictive RAIR findings when 3D high-resolution ARM (HRARM) is used compared to 2D HRARM consecutively on the same patients. 63 However, Keshtgar et al in another recent study reported no significant difference in values of anal resting pressure and volume threshold to elicit RAIR obtained with conventional water perfusion or 3D high-resolution ARM. 64 ARM guidelines for performance and interpretation were revised and published by the American Neurogastroenterology and Motility Society in conjunction with The North American Society of Paediatric Gastroenterology, Hepatology and Nutrition 65 and recently by the British Society of Paediatric Gastroenterology, Hepatology and Nutrition (BSPGHAN).…”
Section: Study Performancementioning
confidence: 93%
“…Some have reported differences in the data obtained with different types of catheters. Two recent prospective studies reported a higher anal resting pressure recorded and conflictive RAIR findings when 3D high‐resolution ARM (HRARM) is used compared to 2D HRARM consecutively on the same patients 63 . However, Keshtgar et al.…”
Section: Motility Studiesmentioning
confidence: 99%
“…High-resolution anorectal manometry (HRAM) was performed for all LARS patients as described in a previous study ( 12 ). In brief, one dose of glycerin enema was administered for bowel preparation 30–60 min before the HRAM examination.…”
Section: Methodsmentioning
confidence: 99%
“…Of note, normal values from manometric parameters may differ according to age, with lower values of resting pressure and thresholds for RAIR in neonates compared with older children (Table 1) (49,50). Moreover, recorded pressure values may depend on the specific type of pressure sensor used during the test (water-perfused vs solid-state) (51,52), the diameter of the catheter (greater diameter may produce higher pressure values obtained by solid-state sensors) (53,54), the protocol used by investigator (duration of the recording, eg, resting pressure over 30 or 60 seconds) or even the software used to process the recorded data (maximum, mean, median, or increment) (55). Therefore, it is important to use normal values obtained with the same equipment and adequate protocol.…”
Section: Normal Valuesmentioning
confidence: 99%