2020
DOI: 10.24293/ijcpml.v27i1.1610
|View full text |Cite
|
Sign up to set email alerts
|

Comparison of Blood Gas Analysis between Benchtop and Handheld Device

Abstract: The use of blood gas analysis is to determine the Acid-base status required to treat patients with emergency conditionssuch as metabolic disorders and respiratory diseases. Benchtop device is commonly used in hospitals to analyze blood gas;however, handheld devices are recently more often used in emergency settings due to its quick and simple process. Thisstudy was performed to compare blood gas analysis results between the i-STAT handheld device and the Nova pHox Ultrabenchtop device that were currently being… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1

Citation Types

0
4
0
1

Year Published

2021
2021
2024
2024

Publication Types

Select...
4

Relationship

0
4

Authors

Journals

citations
Cited by 4 publications
(5 citation statements)
references
References 2 publications
0
4
0
1
Order By: Relevance
“…Method comparison studies have been conducted using human whole blood and serum samples for the Nova pHOx Ultra analyzer and the Nova Critical Care Xpress analyzer or i-STAT analyzer. [22][23][24][25] Good agreement was observed in the first two studies, whereas in the latter two studies, the Ultra direct ISE analyzer read whole blood Na concentration 4.9 and 4.3 mmol/L lower, and whole blood K concentration 0.25 and 0.04 mmol/L lower than indirect ISE autoanalyzers. 24,25 For comparison, the Ultra direct ISE analyzer read whole blood Na concentration 2.5 mmol/L lower, and whole blood K concentration 0.11 mmol/L higher than the indirect ISE Beckman chemistry analyzer in our study.…”
Section: Discussionmentioning
confidence: 72%
See 3 more Smart Citations
“…Method comparison studies have been conducted using human whole blood and serum samples for the Nova pHOx Ultra analyzer and the Nova Critical Care Xpress analyzer or i-STAT analyzer. [22][23][24][25] Good agreement was observed in the first two studies, whereas in the latter two studies, the Ultra direct ISE analyzer read whole blood Na concentration 4.9 and 4.3 mmol/L lower, and whole blood K concentration 0.25 and 0.04 mmol/L lower than indirect ISE autoanalyzers. 24,25 For comparison, the Ultra direct ISE analyzer read whole blood Na concentration 2.5 mmol/L lower, and whole blood K concentration 0.11 mmol/L higher than the indirect ISE Beckman chemistry analyzer in our study.…”
Section: Discussionmentioning
confidence: 72%
“…Finally, samples were not analyzed in duplicate, as recommended. 2 Blood gas chemistry sample analyses from critically ill animals are rarely run in duplicate 10,[22][23][24][25] because of the added cost and time in obtaining results and time-related changes in analytes. Consequently, single measurement was preferred in our study as it reflected the use of the analyzers in veterinary medicine.…”
Section: Discussionmentioning
confidence: 99%
See 2 more Smart Citations
“…15 Hal ini sesuai dengan penelitian lainnya oleh Liana dkk (2019) menggunakan alat i-STAT dan Nova pHox Ultra dengan total 42 sampel terdapat korelasi yang bermakna (p<0,05) dan hubungan sangat kuat (r≥0,8). 16 Sebuah studi lainnya yang dilakukan oleh Boonlert dkk (2003) ditemukan adanya korelasi antara hasil pemeriksaan POCT analisis gas darah OPTI CCA dan OMNI 9 dengan alat analisis gas darah di rumah sakit (SP, RxL, CRT, san Cell Dyn) menunjukan adanya korelasi antara kedua hasil pemeriksaan. 17 Pada penelitian yang dilakukan oleh Lukkonen dkk (2015) menggunakan alat POCT EPOC dengan laboratory blood gas analyzer Rapidlab RL1265 dan Rapid point RP500, ditemukan adanya hubungan antara hasil analisis gas darah menggunakan alat POCT EPOC dengan laboratory blood gas analyzer Rapidlab RL1265 dan Rapid point RP500.…”
Section: Pembahasanunclassified