Abstract:Abstract. A pediatric anthropomorphic phantom was irradiated using a standard 3DCRT treatment plan with and without including the couch in the Treatment Planning System (TPS; Varian Eclipse 13.5, AAA algorithm) calculations. The 2D scatter dose distributions at several planes away from the isocenter were measured using a Gafchromic EBT3 radiochromic film, and compared with the calculated ones at several distances from the isocenter. The out-of-field dose distributions (from TPS and EBT3) differ considerably, p… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.