Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kV to 300 kV

Abstract: (b) , D. Butler (c) , I. Csete (d) , F. Gabris (e) , A. Hakanen (f) , J.-H. Lee (g) , M. Palmer (h) , N. Saito (i) and W. de Vries (j)

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
6
0

Year Published

2009
2009
2019
2019

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 7 publications
(6 citation statements)
references
References 3 publications
0
6
0
Order By: Relevance
“…The comparison results, R, for the selected beam qualities are the mean of the IAEA/MKEH ratios in the final column of Tables 4 and 5 for the two transfer chambers. It is interesting to note that these ratios are close to unity despite the fact that the results for the MKEH and the PTB (where the IAEA reference chamber was calibrated) in the EURAMET.RI(I)-S3 have a systematic difference as seen in column 8 of Table 8 showing the PTB/MKEH ratio R K taken from [3]. Normalizing the present results for IAEA/MKEH to the results for PTB/MKEH obtained in the EURAMET.RI(I)-S3 comparison would give the results R/R K for IAEA/PTB shown in the last two columns of the table.…”
Section: Resultsmentioning
confidence: 89%
See 3 more Smart Citations
“…The comparison results, R, for the selected beam qualities are the mean of the IAEA/MKEH ratios in the final column of Tables 4 and 5 for the two transfer chambers. It is interesting to note that these ratios are close to unity despite the fact that the results for the MKEH and the PTB (where the IAEA reference chamber was calibrated) in the EURAMET.RI(I)-S3 have a systematic difference as seen in column 8 of Table 8 showing the PTB/MKEH ratio R K taken from [3]. Normalizing the present results for IAEA/MKEH to the results for PTB/MKEH obtained in the EURAMET.RI(I)-S3 comparison would give the results R/R K for IAEA/PTB shown in the last two columns of the table.…”
Section: Resultsmentioning
confidence: 89%
“…Details of the MKEH primary standard free-air chamber for air kerma and its uncertainty are given in [3]. For the calibration of the transfer chambers the secondary standard chamber, type ND 1001 (#7807) was used, having a volume of 22 cm 3 and diameter39 mm.…”
Section: Air Kerma Determination At the Mkehmentioning
confidence: 99%
See 2 more Smart Citations
“…The ISO 4037 narrow‐spectrum series X‐ray qualities are mainly used to calibrate radiation protection detectors of very large dimensions. Existing key comparison for air kerma for the ISO 4037 narrow‐spectrum series that cover the range from 30 to 300 kV is not direct comparison, such as BIPM.RI(I)‐K2 and ‐K3, but indirect comparison, such as EUROMET.RI(I)‐S3 whose results are available [3–5].…”
Section: Introductionmentioning
confidence: 99%