Comparison of Signal and Bactec NR‐660 blood culture systems

Abstract: The Signal blood culture system was compared with the Bactec NR-660. A total of 1617 blood culture sets yielded 143 (8.8%) significant isolates; 113 (79.0%) were from positive bottles in both the Bactec and Signal systems. Twelve organisms (8.4%) were detected and isolated from the Signal system only and another 18 (12.6%) from the Bactec system only. Of these 18, five were Signal-positive but the organism was not recovered and four organisms were isolated from negative Signal bottles on terminal subculture. T… Show more