Comparison of total-body measurements by dual-energy x-ray absorptiometry and dual-photon absorptiometry

Abstract: Both dual-photon absorptiometry (DPA) using 153Gd and dual-energy x-ray absorptiometry (DEXA) can be used for measurement of bone mineral content (BMC) and bone mineral density (BMD) of the total skeleton and its seven major regions. The short-term precision (coefficient of variation, CV) of DEXA for total-body BMD using the medium (20 minute) and fast (10 minute) speeds was 0.34 and 0.68% in 5 normal subjects; the corresponding CV in 5 osteoporotic females were 0.70 and 1.04%. The CV for BMD using DPA was 0.8… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

1
2
0

Year Published

1993
1993
2009
2009

Publication Types

Select...
6
2

Relationship

0
8

Authors

Journals

citations
Cited by 36 publications
(3 citation statements)
references
References 24 publications
1
2
0
Order By: Relevance
“…The precision of densitometric parameters was assessed by repeated DEXA measurements in 14 CHF patients (mean duration between assessments 4±1 days, range 2–7). Coefficients of variance for repeated measurements for all parameters were below 3%, which was in accordance with other studies showing a short‐term variability of <5% ,,…”
Section: Methodssupporting
confidence: 92%
“…The precision of densitometric parameters was assessed by repeated DEXA measurements in 14 CHF patients (mean duration between assessments 4±1 days, range 2–7). Coefficients of variance for repeated measurements for all parameters were below 3%, which was in accordance with other studies showing a short‐term variability of <5% ,,…”
Section: Methodssupporting
confidence: 92%
“…DEXA is a widely applied, non-invasive and reproducible method for the assessment of body composition, and involves only a minimal exposure to ionizing radiation [22].…”
Section: Densitometry Measurement Of Lean Tissue Massmentioning
confidence: 99%
“…The T ‐ and Z ‐scores were estimated using mean BMD and SD values supplied by the Hologic equipment manufacturer. The reproducibility of the whole body BMD is calculated as <0.01 g/cm 2 or 0.1 SD (17), (18).…”
Section: Methodsmentioning
confidence: 99%