2023
DOI: 10.1002/lary.31106
|View full text |Cite
|
Sign up to set email alerts
|

Concurrent Validity of a Low‐Cost Manometer for Objective Assessments of Respiratory Muscle Strength

James A. Curtis,
Valentina Mocchetti,
Anaïs Rameau

Abstract: Objective(s)This study examined the agreement in maximal expiratory (MEP) and inspiratory (MIP) pressure readings between two digital manometers: (1) the MicroRPM – the gold‐standard manometer for respiratory muscle strength testing; and (2) the LDM – a low‐cost, commercially available, alternative manometer.MethodsPositive (MEP) and negative (MIP) pressures were simultaneously applied to the MicroRPM and LDM using a 3‐liter syringe within a controlled laboratory setting. Pressure readings were compared, and a… Show more