ObjectivesContrast-enhanced spectral mammography (CESM) is a promising problem-solving tool in women referred from a breast cancer screening program. We aimed to study the validity of preliminary results of CESM using a larger panel of radiologists with different levels of CESM experience.MethodsAll women referred from the Dutch breast cancer screening program were eligible for CESM. 199 consecutive cases were viewed by ten radiologists. Four had extensive CESM experience, three had no CESM experience but were experienced breast radiologists, and three were residents. All readers provided a BI-RADS score for the low-energy CESM images first, after which the score could be adjusted when viewing the entire CESM exam. BI-RADS 1-3 were considered benign and BI-RADS 4-5 malignant. With this cutoff, we calculated sensitivity, specificity and area under the ROC curve.ResultsCESM increased diagnostic accuracy in all readers. The performance for all readers using CESM was: sensitivity 96.9 % (+3.9 %), specificity 69.7 % (+33.8 %) and area under the ROC curve 0.833 (+0.188).ConclusionCESM is superior to conventional mammography, with excellent problem-solving capabilities in women referred from the breast cancer screening program. Previous results were confirmed even in a larger panel of readers with varying CESM experience.Key Points• CESM is consistently superior to conventional mammography
• CESM increases diagnostic accuracy regardless of a reader’s experience
• CESM is an excellent problem-solving tool in recalls from screening programs