Contrast-enhanced Spectral Mammography

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
16
0

Year Published

2018
2018
2024
2024

Publication Types

Select...
5
1

Relationship

0
6

Authors

Journals

citations
Cited by 25 publications
(16 citation statements)
references
References 9 publications
0
16
0
Order By: Relevance
“…Like other imaging modalities, CEM imaging artifacts can occur ( Fig. 4), challenging interpretation [25][26][27]. Rim and skin line artifacts, due to scatter radiation from non-uniform thickness of breast tissue and skin, can mimic enhancement.…”
Section: Cem Interpretation and Reportingmentioning
confidence: 99%
“…Like other imaging modalities, CEM imaging artifacts can occur ( Fig. 4), challenging interpretation [25][26][27]. Rim and skin line artifacts, due to scatter radiation from non-uniform thickness of breast tissue and skin, can mimic enhancement.…”
Section: Cem Interpretation and Reportingmentioning
confidence: 99%
“…Some FN lesions had a copious post biopsy hematoma at the site of the index lesion which generated a characteristic artifact called “negative contrast enhancement” or “eclipse sign”, thus preventing correct visualization of the BC and interpretation of the images [ 28 , 29 ]. In these cases, only the artifact generated by the hematoma was visible with no other area of enhancement near it.…”
Section: Methodsmentioning
confidence: 99%
“…This is a software-processing artefact also known as "breast-within-a-breast artefact" [5]. It occurs due to a rapid difference of scattered radiation between different breast thicknesses separating the central and peripheral regions causing the software-processing algorithm to create a false exaggerated boundary.…”
Section: Halo Artefactmentioning
confidence: 99%
“…Dromain et al [12] suggested that this may be attributed to patient motion, due to the short interval between the low-and high-energy exposures. However, there are some authors [5] who attribute it to cardiac pulsations transmitted through the chest wall as it more commonly seen in the lower quadrants of MLO projections of the left breast. Reducing patient anxiety might reduce this artefact; however, it does not compromise the image quality.…”
Section: Ripple Artefactmentioning
confidence: 99%
See 1 more Smart Citation