Abstract:As a result of experimental work, the linear nature of the temperature drift in the AEMP-14 equipment was established. The temperature coefficient of the resistivity readings depends on the measurement frequency, but does not depend on the signal level. A method for eliminating the temperature drift of the device is proposed. Temperature dependence introduces a deviation from true electrical conductivity within 1–12 % at 20 °C. This method minimizes the error in the measured values.
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.