Introduction: Pulsed-field ablation (PFA), an ablative method that causes cell death by irreversible electroporation, has potential safety advantages over radiofrequency ablation and cryoablation. Pulmonary vein (PV) isolation was performed in a porcine model to characterize safety and performance of a novel, fully-integrated biphasic PFA system comprising a multi-channel generator, variable loop circular catheter, and integrated PFA mapping software module.Methods: Eight healthy porcine subjects were included. To evaluate safety, multiple ablations were performed, including sites not generally targeted for therapeutic ablation, such as the right inferior PV lumen, right superior PV ostium, and adjacent to the esophagus and phrenic nerve. To evaluate the efficacy, animals were recovered, followed for 30(±3) days, then re-mapped. Gross pathological and histopathological examinations assessed procedural injuries, chronic thrombosis, tissue ablation, penetration depth, healing, and inflammatory response.Results: All eight animals survived follow-up. PV narrowing was not observed acutely nor at follow-up, even when ablation was performed deep to the PV ostium.