Cryogenic modulation fast GC × GC–MS using a 10 m microbore column combination: Concept, method optimization, and application

Abstract: The present research is based on the concept of using a 10 m × 0.1 mm id column for cryogenic-modulation fast comprehensive two-dimensional gas chromatography with quadrupole mass spectrometry. Specifically, an 8.9 m × 0.1 mm id low-polarity column was used as the first dimension, and a 1.1 m × 0.1 mm id medium-polarity column was used as the second dimension. The main scope of the investigation was to develop a high peak-capacity method, with an analysis time of approximately 10 min. Various aspects related t… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
9
0

Year Published

2019
2019
2022
2022

Publication Types

Select...
5
1

Relationship

0
6

Authors

Journals

citations
Cited by 8 publications
(9 citation statements)
references
References 17 publications
0
9
0
Order By: Relevance
“…Giocastro et al . analyze a standard mixture of cosmetic allergens in fast‐GC × GC. They also found an increase in signal noise ratio when compared with conventional methods.…”
Section: Resultsmentioning
confidence: 91%
See 2 more Smart Citations
“…Giocastro et al . analyze a standard mixture of cosmetic allergens in fast‐GC × GC. They also found an increase in signal noise ratio when compared with conventional methods.…”
Section: Resultsmentioning
confidence: 91%
“…The high heating rates provided narrower peaks, as a consequence, the signal noise ratio values were increased, showing another peculiarity of fast-GC × GC technique [14,15,17]. Giocastro et al [18] analyze a standard mixture of cosmetic allergens in fast-GC × GC. They also found an increase in signal noise ratio when compared with conventional methods.…”
Section: Standard Mixturementioning
confidence: 99%
See 1 more Smart Citation
“…× 0.1 μm d f 1 D column) connected to a midpolar 1.1 m × 0.1 mm i.d. × 0.1 μm d f 2 D column, using a loop thermal modulator had a total analysis time of 10 min, with satisfactory separation performance. Alternatively, modeling can be used to minimize development cost.…”
Section: Instrumental Advancesmentioning
confidence: 99%
“…Additionally, the GC run is often limited to isothermal conditions or less dynamic temperature programmes because of the time limitations. A unique fastGC approach was recently described with the extension of fast GC to two-dimensional fast GC, substantially increasing separation performance while keeping the total GC runtime below 20 min 8 . The addition of fast gas chromatography with runtimes below 20 min (fastGC) to different mass analysers offers a compromi s e between chemical specificity and time resolution.…”
Section: Introductionmentioning
confidence: 99%