Debonding of Porous Coating: A Late Failure Mode of Uncemented, Partially Threaded Acetabular Components—Retrieval Analysis
Łukasz Łapaj,
Joanna Sulej-Chojnacka,
Justyna Rozwalka
et al.
Abstract:Titanium plasma-sprayed (TPS) porous coatings have been used in total hip arthroplasty for decades. They are considered reliable, and very few failure cases have been described so far. This retrieval study described a series of 20 acetabular components—where total or partial debonding occurred during in vivo use and aimed to explain the underlying failure mechanisms. Implants were examined using optical and electron microscopy (SEM), metallographic sections of retrievals were prepared while pathologic samples … Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.