ObjectiveTo evaluate the diagnostic performance of manual grading of anterior segment optical coherence tomography (AS-OCT) in detection of plateau iris configuration (PIC) based on the presence of standardised ultrasound biomicroscopy (UBM) criteria in at least two quadrants; namely, clinical diagnosis of PIC (DxPIC).Methods and analysisIn this cross-sectional study, paired AS-OCT and UBM images were evaluated by three glaucoma specialists. AS-OCT was classified into two mechanisms, PIC versus non-PIC, of primary angle closure disease (PACD) and AS-OCT-PIC diagnostic performance was tested with DxPIC.ResultsOne hundred and seventy-nine eyes of 142 patients were enrolled for analysis, and DxPIC was found in 85 eyes (47.49%). Intraobserver agreement rates of AS-OCT classification by the graders were 0.77, 0.701 and 0.742 (all p<0.001), and interobserver agreement rates, between a senior glaucoma specialist and the other two glaucoma specialists, were 0.68 and 0.702 (all p<0.001). Plateau iris was classified in AS-OCT images by the three graders, rated 32.96%–39.1% and 24.58%–34.08% in the horizontal and vertical axes, respectively. Diagnostic performance was analysed, yielding sensitivity ranging from 56.47% to 77.78%, and specificity of 48.94% to 64.29%. We applied disease prevalence of 30%, revealing positive predictive values varying from 32.16% to 44.44%, and negative predictive values of 72.4% to 85.71%. Accuracy ranged from 51.2% to 65%. Agreement between the two devices was fair, kappa range 0.31–0.351.ConclusionPerformance of manual grading of AS-OCT in detection of DxPIC was relatively poor; therefore, unadjusted AS-OCT does not appear to be good for manual PIC screening in PACD patients and cannot serve as a substitute for UBM in PIC detection.