Delayed ICD Lead Cardiac Perforation: Comparison of Small versus Standard‐Diameter Leads Implanted in a Single Center

Abstract: Small-diameter active-fixation ICD leads are at increased risk of DCP, a finding mostly due to the higher incidence of events in the Riata family. By contrast, passive-fixation small-diameter leads and standard-diameter leads appear to be safe enough regarding the risk of DCP. Our preliminary data suggest that the new Durata lead is not associated with an increased risk of DCP.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
5

Citation Types

1
20
0

Year Published

2015
2015
2018
2018

Publication Types

Select...
5
1

Relationship

0
6

Authors

Journals

citations
Cited by 26 publications
(21 citation statements)
references
References 21 publications
1
20
0
Order By: Relevance
“…Active fixation small‐diameter ICD leads have recently been reported to be associated with an increased risk of delayed cardiac perforation 2. However, in the present case, a standard diameter (8 French) Riata lead was used.…”
Section: Discussionmentioning
confidence: 91%
See 2 more Smart Citations
“…Active fixation small‐diameter ICD leads have recently been reported to be associated with an increased risk of delayed cardiac perforation 2. However, in the present case, a standard diameter (8 French) Riata lead was used.…”
Section: Discussionmentioning
confidence: 91%
“…The incidence of perforation by Riata leads is reported to range from 0.345% to 3.8%, which is relatively high 1, 2. Perforation usually occurs shortly after implantation; cases occurring after 1 month are considered to be rare.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…Small-diameter (≤8 Fr), active-fixation leads were associated with a higher risk of cardiac perforation, in 1.6% of patients, and active-fixation leads in 1.4% 9. Female sex, age >80 years, lead location, lead stiffness and calibre size are risk factors for cardiac lead perforation 10…”
Section: Discussionmentioning
confidence: 99%
“…To date, pediatric patient size has not been reported as a risk factor, but knowledge about this population is limited. Subacute perforations occur more commonly with active fixation leads and those with small diameters, particularly leads in the Riata (St. Jude Medical) family 2, 3. To date there has been only 1 report of cardiac perforation due to a Durata lead 4 .…”
Section: Discussionmentioning