“…Before each run, the instrument was tuned for maximum sensitivity and signal stability to minimize the formation of oxides and double-charge ions. The main operational parameters of the ICP-MS were: nebulizer gas flow, 1.14 L/min; auxiliary gas flow, 0.79 L/min; plasma gas flow, 13.9 L/min; radiofrequency generator power, 1550 W; and dwell time, 1–10 m. The isotopes 7Li, 25Mg, 27Al, 31P, 43Ca, 52Cr, 55Mn, 57Fe, 59Co, 60Ni, 65Cu, 66Zn, 75As, 82Se, 85Rb, 88Sr, 111Cd, 121Sb, 133Cs, 137Ba, 205 TL and 208 Pb were measured for analytical determination and the isotopes 6Li, 45Sc, 71Ga, 89Y, 103Rh, 193Ir and 209Bi were monitored as IS ( Azevedo et al, 2023 ).…”