This work aimed to establish an HILIC‐MS/MS method to simultaneously determine the levels of 13 endogenous amino acids and trimethylamine oxide in the biological samples from the mice. Electrospray ion source was used for the analysis of mass spectrometry. The 20 min separation was applied in a Dikma Inspire Hilic column (2.1 × 100.0 mm, 3 μM). Positive ion mode under an MRM model gave a satisfying response value. The limits of quantitation were evaluated by accuracy from −12.59% to 7.89% and precision from 1.77% to 14.00% as well as acceptable interday and intraday precision, matrix effect, recovery, and stability. Later, the assay was successfully used to measure the concentrations of the determinands in the biological samples. Individual and tissue distribution differences for these metabolites were observable. The amino acids had a consistent highest content in the spleens, while the lowest levels were found in the livers. Alanine was the most abundant amino acid in the serum, and taurine kept the highest content in all of the tissues. Trimethylamine oxide remained low level, especially in the liver samples.