Determination of cortisol cut-off limits and steroid dynamics in ACTH stimulation (Synacthen®) test: A comparative analysis using Roche Elecsys Cortisol II immunoassay and LC-MS/MS
Sema Okutan,
Nanna Thurmann Jørgensen,
Lars Engers Pedersen
et al.
Abstract:Purpose
Measurement of cortisol concentrations is method dependent. Therefore, assay-specific cut-off limits for the definition of e.g., adrenal insufficiency should be available. The study aimed to establish assay-specific cut-off limits for cortisol after adrenocorticotropic hormone (ACTH) stimulation, comparing Roche Elecsys Cortisol II immunoassay to liquid chromatography-mass spectrometry (LC-MS/MS).
Methods
One hundred healthy participants underwent a ACTH test (250 µg), with plasma samples analyzed us… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.