2017
DOI: 10.1088/1748-0221/12/07/t07006
|View full text |Cite
|
Sign up to set email alerts
|

Determination of efficiency of an aged HPGe detector for gaseous sources by self absorption correction and point source methods

Abstract: A: Methods for the determination of efficiency of an aged high purity germanium (HPGe) detector for gaseous sources have been presented in the paper. X-ray radiography of the detector has been performed to get detector dimensions for computational purposes. The dead layer thickness of HPGe detector has been ascertained from experiments and Monte Carlo computations. Experimental work with standard point and liquid sources in several cylindrical geometries has been undertaken for obtaining energy dependant effic… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2019
2019
2023
2023

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 6 publications
(1 citation statement)
references
References 22 publications
0
1
0
Order By: Relevance
“…Huy et al could trace an efficiency reduction back to an increase in the lithium dead layer of their ptype coaxial detector [46]. See also [47,48] and references therein.…”
Section: Dead Regions and Possible Aging Effectsmentioning
confidence: 99%
“…Huy et al could trace an efficiency reduction back to an increase in the lithium dead layer of their ptype coaxial detector [46]. See also [47,48] and references therein.…”
Section: Dead Regions and Possible Aging Effectsmentioning