2023
DOI: 10.22159/ijpps.2023v15i9.48431
|View full text |Cite
|
Sign up to set email alerts
|

Development and Validation of Novel Uv-Visible Spectrophotometric Method for Estimation of Tepotinib in Bulk and in Pharmaceutical Formulation

MONIKA SHIRWAR,
SHUBHANGI BIRAJDAR,
SHRUTI GARAD
et al.

Abstract: Objective: The objective of the study is to develop a new, simple, rapid, accurate, and economical UV-spectrophotometric method and validate the same for the estimation of Tepotinib in bulk and in pharmaceutical formulation as per ICH guidelines. Methods: The present work was carried out using the UV-Visible double-beam spectrophotometer model Systronics 2201. Tepotinib was found soluble in water, methanol, methanol so, analytical-grade methanol was used as a solvent for conducting the work. The λmax of the te… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2

Citation Types

0
2
0

Year Published

2024
2024
2024
2024

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(2 citation statements)
references
References 15 publications
0
2
0
Order By: Relevance
“…The absorbance was measured for the same concentration solution of 5 μg/ml six times. The results as shown in table 5, fell within the acceptable range, with a % RSD of less than 2% [16,17]. The result suggests that the UV method demonstrates good ruggedness, indicating its reliability when implemented by different analysts [21].…”
Section: Ruggednessmentioning
confidence: 89%
See 1 more Smart Citation
“…The absorbance was measured for the same concentration solution of 5 μg/ml six times. The results as shown in table 5, fell within the acceptable range, with a % RSD of less than 2% [16,17]. The result suggests that the UV method demonstrates good ruggedness, indicating its reliability when implemented by different analysts [21].…”
Section: Ruggednessmentioning
confidence: 89%
“…The ruggedness of the proposed method was assessed for a concentration of 5 μg/ml of Linezolid by analysing aliquots from a homogenous slot. Two analysts performed the analysis under identical operational and environmental conditions [16,17].…”
Section: Ruggednessmentioning
confidence: 99%